Просмотр инструкции измерительного прибора Condtrol DeuX, страница 41

Condtrol
41
CONDTROL DeuX/UniX
2
/UniX
3
/UniX
5
User Manual
CROSS LINE
LASER
The difference L between point III and the centre of the laser line at the upper
edge of the door opening results in the actual deviation of the instrument
from the vertical plane.
Measure the height of the door opening.
The maximum permitted deviation L is calculated as follows:
L = double height of the door opening x 0.2 mm/m
Example: With a door opening height of 2 meters, the maximum permitted
deviation is
L = 2 x 2 m x 0.2 mm/m = 0.8 mm,
thus, the marks must not be more than 0.8 mm apart.
When the maximum deviation exceeds 0.8 mm, you should contact
CONDTROL service centre.
Attention!
Always use only the middle of the laser line for marks. The width of the laser
line changes with distance. The instrument automatically switches off in 60
minutes of continuous work. If it is necessary, switch on the instrument and
continue working.
Orthogonal directions construction (UniX)
Check the 90° including angle of cross dot and the front, left, right dots.
Installing the instrument on the tripod, and put them in the centre of 6m x 6m
room, then switch on all the laser beams and dots.
Face the cross dot and front, left, right dots to the walls, the down dot aims
the floor, mark A, B, C, D, E respectively.
Rotate the instrument 90° around the down dot, mark A1, B1, C1, D1 on the
walls, measure respectively the distances of A to C1, B to D1, C to B1 and
D to A1.
If the value is larger than 2 mm, the instrument must be returned to
CONDTROL service centre.