CONDTROL DeuX/UniX
2
/UniX
3
/UniX
5
User Manual
CROSS LINE
LASER
Align the height of the instrument (using a tripod or by underlaying, if
required) in such a manner that the cross point of the laser lines is projected
against the previously marked point II on the wall B.
Without changing the height, turn around the instrument by 180°. Direct it
against the wall A such a manner that the vertical laser line runs through the
already marked point I. Allow the instrument to level in and mark the cross
point of the laser lines on the wall A (point III).
The difference L between both marked points I and III on the wall A indicates
the actual height deviation of the instrument.
The maximum permitted deviation L is calculated as follows:
L = double distance between the walls x 0.2 mm/m
Example: With a 5 meter distance between the walls the maximum deviation is
L = 2 x 5m x 0.2 mm/m = 2 mm,
thus, the marks must not be more than 2 mm apart.
When the maximum deviation exceeds 2 mm, you should contact CONDTROL
service centre.
Horizontal line tilt accuracy
A free area approximately 5*5 m is necessary in order to make control:
-place the instrument on a solid, flat ground in the middle between the walls
A and B. Wait for the ending of compensation in the horizontal mode.
fig. 1