4.Cleaning and disinfecting
o: applicable X: not applicable
A. Use a commercially available disinfectant. Follow the instructions provided by the
disinfectant manufacturer.
Note: Never clean with benzene, thinner or a flammable chemical.
B. Parts may be boiled between 15 to 20 minutes.
After boiling, carefully remove the parts, shake off excess water
and allow to air dry in a clean environment.
Disinfecting
method
Parts
Mouthpiece
Nebulizer Kit
Air Tube
(Silicone, 100cm)
(optional)
Infant Mask (PVC)
Child Mask (PVC)
Kids Accessory
Air Tube
(PVC, 100cm)
Air Filter
Materials
PP
PP
(Vaporizer Head: PC)
Tube: Silicone
Connector: PP
Mask: PVC
(Phthalate free)
Band: Rubber***
PP, PVC
(Phthalate free)
PVC
(Phthalate free)
Polyester
A
Alcohol
oo o o**
XXX
Disinfecting
ethanol
Sodium
hypochlorite
oo o o** o
XX
Milton*
(0.1%, 15min.)
Quaternary
ammonium
oo o o** o
XX
Osvan*
(0.1%, 10min.)
Chlorhexidine
oo o o** o
XX
Hibitane*
(0.5%, 30min.)
Amphoteric
Surfactant
oo o o** o
XX
Tego 51*
(0.2%, 15min.)
B Boiling
oo o
XXXX
* an example of commercially available disinfectant. (The concentration and residence time
specified in the table are under the conditions where the service life of the parts is tested with
each disinfectant used as described in its instruction manual. Please note that the testing was
not performed for the purpose of ensuring the effectiveness of the disinfectants. There is no
intention of suggesting that such disinfectants be used. Conditions of use and ingredients of
disinfectant vary for manufacturers.
Please read the instruction manuals carefully before use and apply disinfectant to each part in
an appropriate way. Please notice that service life of the parts may be shorter depending on
conditions, environments and frequency of use.)
** Remove the rubber band from mask before disinfecting.
***The band is not made with natural rubber latex.